FM Khan, GD Fullerton, JM Lee, VC Moore and SH Levitt
The effect of different parameters on dose distribution in electron- beam arc therapy was studied in order to develop a technique for routine clinical use. A special diaphragm was designed to facilitate telecentric rotation. Dosimetry was performed with an ion chamber, film, and LiF powder in cylindrical polystyrene phantoms and an Alderson Rando phantom. Dose distributions were evaluated with regard to dose homogeneity, and a method of sharpening the dose fall-off near the ends of the arc was proposed. Criteria for selection of isocenter depth and field size were developed. Methods of dose calculation, calibration, and treatment planning are discussed.
This article has been cited by other articles:
![]() |
I. Aubineau-Laniece, L. de Carlan, I. Clairand, A. Lemosquet, S. Chiavassa, N. Pierrat, M. Bardies, and D. Franck Current developments at IRSN on computational tools dedicated to assessing doses for both internal and external exposure Radiat Prot Dosimetry, December 20, 2005; 115(1-4): 522 - 529. [Abstract] [Full Text] [PDF] |
||||