Radiology, Vol 178, 643-646, Copyright © 1991 by Radiological Society of North America
ARTICLES |
CH Cagnon, SH Benedict, NJ Mankovich, JT Bushberg, JA Seibert and JS Whiting
Department of Radiological Sciences, University of California, Los Angeles.
High-level fluoroscopic boost options that exceed conventional exposure limits are available as a means of reducing quantum mottle during angiography. Federal law does not specify exposure limits for such high- level controls but requires specific means of activation to safeguard against inadvertent use. The American Association of Physicists in Medicine recently recommended that high-level exposure rates not exceed 2.58 mC/kg/min (10R/min). At six institution surveyed, maximum exposure rates ranged from 5.42 to 24 mC/kg/min (21-93 R/min). Activation of high-level capability varied from a simple foot switch to a keyed interlock requiring a second operator to engage. There appears to be no industry coherence in high-level control exposure limits as yet, although the Center for Devices and Radiological Health recently initiated an investigatory program.
This article has been cited by other articles:
![]() |
E Kuon, C Glaser, and J B Dahm Effective techniques for reduction of radiation dosage to patients undergoing invasive cardiac procedures Br. J. Radiol., June 1, 2003; 76(906): 406 - 413. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Berlin Radiation-Induced Skin Injuries and Fluoroscopy Am. J. Roentgenol., July 1, 2001; 177(1): 21 - 25. [Full Text] [PDF] |
||||
![]() |
J. T. Cusma, M. R. Bell, M. A. Wondrow, J. P. Taubel, and D. R. Holmes Jr. Real-time measurement of radiation exposure to patients during diagnostic coronary angiography and percutaneous interventional procedures J. Am. Coll. Cardiol., February 1, 1999; 33(2): 427 - 435. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |