Radiology
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Rudin, S
Right arrow Articles by Miller, J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Rudin, S
Right arrow Articles by Miller, J.

Radiology, Vol 178, 647-651, Copyright © 1991 by Radiological Society of North America


ARTICLES

Dose reduction during fluoroscopic placement of feeding tubes

S Rudin, DR Bednarek and JA Miller
Department of Radiology, School of Medicine and Biomedical Sciences, State University of New York, Buffalo 14215.

By both increasing the optical iris of the video camera and removing the grid in fluoroscopic procedures involved in placement of a Dubbhoff feeding tube, the radiation dose to patients and staff was reduced by five to seven times. An average expected dose to a patient of about 300 mGy per procedure was reduced by two to three times when the grid was removed and by an additional three times when an iris of increased diameter was used. Because Dobbhoff procedures do not involve obtaining a diagnosis, the image degradation was acceptable and did not affect the total exposure times or ability to conduct the procedure. With the grid out, the difference between the mean exposure times of 5.1 minutes in 96 patient studies done with a normal iris opening and 4.0 minutes in 52 studies done with an enlarged iris was not statistically significant. The importance of reducing patient dose is reinforced by the finding that one-third of the patients underwent repeated procedures, accounting for almost 60% of the total.


This article has been cited by other articles:


Home page
RadiologyHome page
T. Xu, H. Q. Le, and S. Molloi
Patient-specific Region-of-Interest Fluoroscopy Device for X-ray Dose Reduction
Radiology, February 1, 2003; 226(2): 585 - 592.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
RADIOLOGY RADIOGRAPHICS RSNA JOURNALS ONLINE
Copyright © 1991 by the Radiological Society of North America.