Radiology, Vol 178, 653-658, Copyright © 1991 by Radiological Society of North America
ARTICLES |
OH Suleiman, J Anderson, B Jones, GU Rao and M Rosenstein
Office of Training and Assistance, Food and Drug Administration, Rockville, Md. 20857.
A method was developed to estimate tissue doses from the upper gastrointestinal fluoroscopy examination. It involved measuring the technical parameters of the clinical examination, partitioning the dynamic examination into a set of discrete x-ray fields, and generating corresponding tissue does tables with an existing computer program. Knowledge of the radiation exposures associated with each of the fields enabled the calculation of tissue doses for the entire dynamic examination. In this limited sample (eight patients), fluoroscopy times ranged from 108 to 183 seconds. Radiation exposures ranged from 2.3 to 7.2 mC/kg (9.1-28 R), thyroid doses from 0.15 to 3.5 mGy (15-350 mrad), uterine doses from 0.16 to 1.0 mGy (16-100 mrad), lung doses from 0.90 to 4.2 mGy (90-420 mrad), and active bone marrow doses from 0.81 to 5.4 mGy (81-540 mrad).
This article has been cited by other articles:
![]() |
T. A. Jaffe, A. M. Gaca, S. Delaney, T. T. Yoshizumi, G. Toncheva, G. Nguyen, and D. P. Frush Radiation Doses from Small-Bowel Follow-Through and Abdominopelvic MDCT in Crohn's Disease Am. J. Roentgenol., November 1, 2007; 189(5): 1015 - 1022. [Abstract] [Full Text] [PDF] |
||||
![]() |
E Yakoumakis, I A Tsalafoutas, D Nikolaou, I Nazos, E Koulentianos, and C. Proukakis Differences in effective dose estimation from dose-area product and entrance surface dose measurements in intravenous urography Br. J. Radiol., August 1, 2001; 74(884): 727 - 734. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |