Radiology, Vol 181, 661-668, Copyright © 1991 by Radiological Society of North America
ARTICLES |
L Arrive, Y Menu, I Dessarts, B Dubray, MP Vullierme, V Vilgrain, D Najmark and H Nahum
Department of Radiology, Hopital Beaujon, Clichy, France.
The authors compared the accuracy of spin-echo (SE) and gradient-echo (GRE) magnetic resonance images in diagnosis of abdominal venous thrombosis. Images of 292 abdominal veins in 72 patients were reviewed in a blinded manner by three radiologists, with seven levels of confidence for diagnosis. Corroborative studies proved thrombosis (n = 95) and vessel patency (n = 197). Receiver operating characteristic (ROC) curves were constructed for SE images alone, GRE images alone, and SE and GRE images combined. At specificities of 90% and 95%, thrombi were identified with sensitivities, respectively, of 76% and 63% with SE images, 74% and 58% with GRE images, and 88% and 82% with combined SE and GRE images. The area under the ROC curve for SE and GRE images combined (0.958 +/- 0.011 [standard deviation]) was significantly larger (P less than .001) than that for SE images alone (0.913 +/- 0.018) and GRE images alone (0.921 +/- 0.016). It is concluded that combination of SE and GRE images significantly increases the accuracy of diagnosis of abdominal venous thrombosis.
This article has been cited by other articles:
![]() |
J. Y. Jeong, S. H. Kim, H. J. Lee, and J. S. Sim Atypical Low-Signal-Intensity Renal Parenchyma: Causes and Patterns RadioGraphics, July 1, 2002; 22(4): 833 - 846. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. B. Rafal, W. M. Williams, and J. A. Markisz Venous Thrombosis: Detection via Magnetic Resonance Imaging Vascular and Endovascular Surgery, March 1, 1993; 27(2): 115 - 121. [Abstract] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |