1 From the Department of Diagnostic Radiology, Universitätsspital Zürich, Rämistrasse 100, CH-8091 Zürich, Switzerland (T.B.); Department of Diagnostic and Interventional Radiology, Friedrich-Schiller-Universität Jena, Germany (A.M., J.R.R., I.H., M.F., W.A.K.); Department of Radiology, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, Mass (S.N.G.); and Schering AG, Research Laboratories, Berlin, Germany (P.H., M.R.). Received March 7, 2001; revision requested April 9; revision received June 18; accepted July 24. Address correspondence to T.B. (e-mail: thomas_boehm@gmx.net).
PURPOSE: To compare two methods of radio-frequency (RF) ablation, saline enhancement technique and internally cooled electrodes, for the treatment of small breast cancers in an animal modelhighly aggressive VX2 rabbit tumors surrounded by adipose tissue.
MATERIALS AND METHODS: Twenty-seven tumors were implanted into retroperitoneal fat of 14 New Zealand White rabbits. RF ablation was performed with ultrasonographic (US) guidance after tumors had grown to 15 mm. Fourteen tumors in seven animals were treated with internally cooled electrodes (30-mm-tip single electrode, 60 W, 10 min); 13 tumors in seven animals, with saline enhancement (0.5 mL/min of saline, 25-mm tip, 30 W, 10 min). Autopsy and histopathologic assessment were performed 3 weeks after therapy.
RESULTS: Real-time US of RF ablation was not possible with either method because of obscuration by the increasing hyperechogenicity of the tumor and the surrounding adipose tissue. Equivalent efficacy was demonstrated with the two methods. Significantly greater complications were observed with the saline technique: Free retroperitoneal fluid was detected in one of seven animals with internally cooled electrodes and in all seven animals with saline enhancement (P < .01). Damage to remote structures such as the kidney, spine muscle, and skin was observed at autopsy in one of seven animals with internally cooled technique versus five of seven with saline enhancement (P < .01).
CONCLUSION: Given a lower complication rate and similar treatment efficacy in an animal tumor model, internally cooled RF electrode may be advantageous to adjuvant saline infusion for the minimally invasive treatment of breast tumors.
© RSNA, 2002
Index terms: Animals Breast neoplasms, 00.30 Radiofrequency (RF) ablation, 00.1269 Ultrasound (US), guidance, 00.12986
This article has been cited by other articles:
![]() |
P. F. Laeseke, L. A. Sampson, T. C. Winter III, and F. T. Lee Jr. Use of Dextrose 5% in Water Instead of Saline to Protect Against Inadvertent Radiofrequency Injuries Am. J. Roentgenol., March 1, 2005; 184(3): 1026 - 1027. [Full Text] [PDF] |
||||
![]() |
J. M. Lee, J. K. Han, S. H. Kim, S. H. Choi, S. K. An, C. J. Han, and B. I. Choi Bipolar Radiofrequency Ablation Using Wet-Cooled Electrodes: An In Vitro Experimental Study in Bovine Liver Am. J. Roentgenol., February 1, 2005; 184(2): 391 - 397. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. L. Pereira, J. Trubenbach, M. Schenk, J. Subke, S. Kroeber, I. Schaefer, C. T. Remy, D. Schmidt, J. Brieger, and C. D. Claussen Radiofrequency Ablation: In Vivo Comparison of Four Commercially Available Devices in Pig Livers Radiology, August 1, 2004; 232(2): 482 - 490. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. A. Shock, K. Meredith, T. F. Warner, L. A. Sampson, A. S. Wright, T. C. Winter III, D. M. Mahvi, J. P. Fine, and F. T. Lee Jr Microwave Ablation with Loop Antenna: In Vivo Porcine Liver Model Radiology, April 1, 2004; 231(1): 143 - 149. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Giorgio, L. Tarantino, G. de Stefano, V. Scala, G. Liorre, F. Scarano, A. Perrotta, N. Farella, V. Aloisio, N. Mariniello, et al. Percutaneous Sonographically Guided Saline-Enhanced Radiofrequency Ablation of Hepatocellular Carcinoma Am. J. Roentgenol., August 1, 2003; 181(2): 479 - 484. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Kettenbach, W. Kostler, E. Rucklinger, B. Gustorff, M. Hupfl, F. Wolf, K. Peer, M. Weigner, J. Lammer, W. Muller, et al. Percutaneous Saline-Enhanced Radiofrequency Ablation of Unresectable Hepatic Tumors: Initial Experience in 26 Patients Am. J. Roentgenol., June 1, 2003; 180(6): 1537 - 1545. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Boehm, A. Malich, S. N. Goldberg, P. Hauff, M. Reinhardt, J. R. Reichenbach, W. Muller, M. Fleck, B. Seifert, and W. A. Kaiser Radio-frequency Ablation of VX2 Rabbit Tumors: Assessment of Completeness of Treatment by Using Contrast-enhanced Harmonic Power Doppler US Radiology, December 1, 2002; 225(3): 815 - 821. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |