1 From the Departments of Diagnostic Radiology (A.F.K., M.H., A.K., C.D.C.) and Cardiology (S.S.), Eberhard-Karls-University Tübingen, Hoppe-Seyler-Strasse 3, 72076 Tübingen, Germany; CT Division, Siemens Medical Engineering, Forchheim, Germany (B.O.); Department of Clinical Radiology, Klinikum Grosshadern, University of Munich, Germany (C.B.); and Department of Radiology, Mayo Clinic, Jacksonville, Fla (R.K.). From the 2000 RSNA scientific assembly. Received December 21, 2000; revision requested January 26, 2001; final revision received March 25, 2002; accepted April 9. Address correspondence to A.F.K. (e-mail: andreas.kopp@uni-tuebingen.de).
PURPOSE: To methodically evaluate the reproducibility and accuracy of coronary arterial calcification measurements by using spiral multidetector row and electron-beam computed tomography (CT) with a beating heart phantom.
MATERIALS AND METHODS: A phantom was built to mimic a beating heart with coronary arteries and calcified plaques. The simulated vessels moved in a pattern similar to that of a beating heart. The phantom operated at a variety of pulse rates (0140 beats per minute). The phantom was repeatedly scanned in various positions by using various protocols with electron-beam and multidetector row CT scanners to assess interexamination variability. Statistical analysis was performed to determine significant differences in interexamination variability for various acquisition protocols.
RESULTS: Electrocardiographically (EKG) gated volume coverage with spiral multidetector row CT (2.5-mm collimation) and overlapping image reconstruction (1-mm increment) was found to significantly improve the reliability of coronary arterial calcium quantification, especially for small plaques (P < .05). Mean interexamination variability was reduced from 35% ± 6 (SD) (Agatston score, standard electron-beam CT) to 4% ± 2 (P < .05) (volumetric score, spiral EKG-gated multidetector row CT).
CONCLUSION: By coupling retrospective gating with nearly isotropic volumetric imaging data, spiral multidetector row CT provides better input data for quantification of coronary arterial calcium volume. Multidetector row CT allows precise and repeated measurement of coronary arterial calcification, with low interexamination variability.
© RSNA, 2002
Index terms: Experimental study Heart, calcification, 51.81 Heart, CT, 51.12115, 51.12119 Phantoms
This article has been cited by other articles:
![]() |
K. A. Barraclough, L. A. Stevens, L. Er, D. Rosenbaum, J. Brown, P. Tiwari, and A. Levin Coronary artery calcification scores in patients with chronic kidney disease prior to dialysis: reliability as a trial outcome measure Nephrol. Dial. Transplant., October 1, 2008; 23(10): 3199 - 3205. [Abstract] [Full Text] [PDF] |
||||
![]() |
M.-T. Wu, P. Yang, Y.-L. Huang, J.-S. Chen, C.-C. Chuo, C. Yeh, and R.-S. Chang Coronary Arterial Calcification on Low-Dose Ungated MDCT for Lung Cancer Screening: Concordance Study with Dedicated Cardiac CT Am. J. Roentgenol., April 1, 2008; 190(4): 923 - 928. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Rutten, I. Isgum, and M. Prokop Coronary Calcification: Effect of Small Variation of Scan Starting Position on Agatston, Volume, and Mass Scores Radiology, December 1, 2007; 246(1): 90 - 98. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Horiguchi, H. Yamamoto, N. Hirai, Y. Akiyama, C. Fujioka, K. Marukawa, H. Fukuda, and K. Ito Variability of repeated coronary artery calcium measurements on low-dose ECG-gated 16-MDCT. Am. J. Roentgenol., July 1, 2006; 187(1): W1 - W6. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. R. Coles, M. A. Smail, I. S. Negus, P. Wilde, M. Oberhoff, K. R. Karsch, and A. Baumbach Comparison of Radiation Doses From Multislice Computed Tomography Coronary Angiography and Conventional Diagnostic Angiography J. Am. Coll. Cardiol., May 2, 2006; 47(9): 1840 - 1845. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. E. Clouse, J. Chen, H. M. Krumholz, M. E. Clouse, J. Chen, and H. M. Krumholz Noninvasive Screening for Coronary Artery Disease With Computed Tomography Is Useful Circulation, January 3, 2006; 113(1): 125 - 146. [Full Text] [PDF] |
||||
![]() |
J. Horiguchi, Y. Shen, Y. Akiyama, N. Hirai, K. Sasaki, M. Ishifuro, T. Nakanishi, and K. Ito Electron Beam CT Versus 16-MDCT on the Variability of Repeated Coronary Artery Calcium Measurements in a Variable Heart Rate Phantom Am. J. Roentgenol., October 1, 2005; 185(4): 995 - 1000. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Horiguchi, H. Yamamoto, Y. Akiyama, N. Hirai, K. Marukawa, H. Fukuda, and K. Ito Variability of Repeated Coronary Artery Calcium Measurements by 16-MDCT with Retrospective Reconstruction Am. J. Roentgenol., June 1, 2005; 184(6): 1917 - 1923. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. S. Halliburton, A. E. Stillman, M. Lieber, J. M. Kasper, S. A. Kuzmiak, and R. D. White Potential Clinical Impact of Variability in the Measurement of Coronary Artery Calcification with Sequential MDCT Am. J. Roentgenol., February 1, 2005; 184(2): 643 - 648. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. H. Jung, K. R. Ma, and H. Han Elevated concentrations of cardiac troponins are associated with severe coronary artery calcification in asymptomatic haemodialysis patients Nephrol. Dial. Transplant., December 1, 2004; 19(12): 3117 - 3123. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. J. Schoepf, C. R. Becker, B. M. Ohnesorge, and E. K. Yucel CT of Coronary Artery Disease Radiology, July 1, 2004; 232(1): 18 - 37. [Abstract] [Full Text] [PDF] |
||||
![]() |
G J Morgan-Hughes, P E Owens, C A Roobottom, and A J Marshall Three dimensional volume quantification of aortic valve calcification using multislice computed tomography Heart, October 1, 2003; 89(10): 1191 - 1194. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. H. Mahnken, J. E. Wildberger, J. Simon, R. Koos, T. G. Flohr, S. Schaller, and R. W. Gunther Detection of Coronary Calcifications: Feasibility of Dose Reduction with a Body Weight-Adapted Examination Protocol Am. J. Roentgenol., August 1, 2003; 181(2): 533 - 538. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. M. Moe, K. D. O'Neill, N. Fineberg, S. Persohn, S. Ahmed, P. Garrett, and C. A. Meyer Assessment of vascular calcification in ESRD patients using spiral CT Nephrol. Dial. Transplant., June 1, 2003; 18(6): 1152 - 1158. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Hecht, J. A. Rumberger, M. J. Budoff, R. Park, R. Detrano, P. Fu, Y. Ibrahim, M. Xiang, L. LaBree, and S. Azen C-Reactive Protein and Electron Beam Tomography * Reply Circulation, May 13, 2003; 107 (18): e123 - e124. [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |