Cardiac Imaging |
1 From the Mallinckrodt Institute of Radiology, Washington University, School of Medicine, 510 S Kingshighway Blvd, St Louis, MO 63110. From the 2001 RSNA Scientific Assembly. Received December 13, 2001; revision requested February 18, 2002; final revision received October 30; accepted November 11. Address correspondence to K.T.B. (e-mail: baet@mir.wustl.edu).
PURPOSE: To assess the interscan variability of coronary artery calcium as measured with different tube currents and quantification methods in prospective electrocardiography (ECG)-gated multidetector row CT.
MATERIALS AND METHODS: Thirty-three subjects who were asymptomatic for coronary heart disease underwent prospective ECG-gated, subsecond multidetector row CT of the heart. Each subject underwent two consecutive CT examinations, the first with a dose of either 40 mAs (n = 18) or 80 mAs (n = 15) and the second with a dose of 150 mAs. Calcium volume and calcium score were calculated. Pearson correlation coefficient was computed between the calcium scores of high- and low-dose examinations. Interscan variability in these measurements (ie, the absolute percentage difference) was compared between the examinations with 40150 mAs and those with 80150 mAs by using an independent sample t test. In addition, the interscan variabilities of calcium scores between vessels were evaluated with repeated measures of analysis of variance. The interscan variabilities between calcium score and volume measurement were also compared with paired t tests.
RESULTS: Twenty-seven of 33 subjects had coronary artery calcium deposits on both CT scans. Five subjects had no calcium deposit on either scan. One subject had calcium deposits on only one scan. The total calcium score between the high- and low-dose scans was highly correlated (r = 0.98) and was not significantly different (P = .58). The interscan variability of calcium score showed no significant difference with respect to subject (P = .25) or vessel (P = .84). The interscan variability of the calcium volume measurement was significantly lower than that of the calcium score with respect to both the subject (P < .01) and the vessel (P < .01).
CONCLUSION: A dose of 40 mAs appears adequate for quantifying coronary artery calcium at multidetector row CT. Interscan variability of multidetector row CT is substantially reduced by using the calcium volume method.
© RSNA, 2003
Index terms: Computed tomography (CT), multidetector row Coronary vessels, calcification, 54.76 Coronary vessels, CT, 949.1291, 54.12119
This article has been cited by other articles:
![]() |
S. M. Kim, M. J. Chung, K. S. Lee, Y. H. Choe, C. A. Yi, and B.-K. Choe Coronary Calcium Screening Using Low-Dose Lung Cancer Screening: Effectiveness of MDCT with Retrospective Reconstruction Am. J. Roentgenol., April 1, 2008; 190(4): 917 - 922. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Budoff, S. Achenbach, R. S. Blumenthal, J. J. Carr, J. G. Goldin, P. Greenland, A. D. Guerci, J. A.C. Lima, D. J. Rader, G. D. Rubin, et al. Assessment of Coronary Artery Disease by Cardiac Computed Tomography: A Scientific Statement From the American Heart Association Committee on Cardiovascular Imaging and Intervention, Council on Cardiovascular Radiology and Intervention, and Committee on Cardiac Imaging, Council on Clinical Cardiology Circulation, October 17, 2006; 114(16): 1761 - 1791. [Full Text] [PDF] |
||||
![]() |
J. Horiguchi, H. Yamamoto, N. Hirai, Y. Akiyama, C. Fujioka, K. Marukawa, H. Fukuda, and K. Ito Variability of repeated coronary artery calcium measurements on low-dose ECG-gated 16-MDCT. Am. J. Roentgenol., July 1, 2006; 187(1): W1 - W6. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Shemesh, R. Evron, N. Koren-Morag, S. Apter, J. Rozenman, D. Shaham, Y. Itzchak, and M. Motro Coronary Artery Calcium Measurement with Multi-Detector Row CT and Low Radiation Dose: Comparison between 55 and 165 mAs Radiology, September 1, 2005; 236(3): 810 - 814. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Hong, T. K. Pilgram, F. Zhu, and K. T. Bae Coronary Artery Calcification: Effect of Size of Field of View on Multi-Detector Row CT Measurements Radiology, October 1, 2004; 233(1): 281 - 285. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |