Radiology
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Published online before print October 19, 2005, 10.1148/radiol.2373041439
This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
2373041439v1
237/3/872    most recent
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Doyle, A. J.
Right arrow Articles by Anderson, G. D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Doyle, A. J.
Right arrow Articles by Anderson, G. D.
(Radiology 2005;237:872-877.)
© RSNA, 2005


Computer Applications

Personal Computer versus Workstation Display: Observer Performance in Detection of Wrist Fractures on Digital Radiographs1

Anthony J. Doyle, MB, ChB, James Le Fevre, BHB and Graeme D. Anderson, MB, ChB

1 From the Radiology Department, Middlemore Hospital, Hospital Rd, Otahuhu, Auckland 6, New Zealand. Received August 18, 2004; revision requested October 21; revision received December 17; accepted January 20, 2005. Address correspondence to A.J.D. (e-mail: Adoyle{at}middlemore.co.nz).

PURPOSE: To retrospectively compare the accuracy of observer performance with personal computer (PC) compared with that with dedicated picture archiving and communication system (PACS) workstation display in the detection of wrist fractures on computed radiographs.

MATERIALS AND METHODS: This study was conducted according to the principles of the Declaration of Helsinki (2002 version) of the World Medical Association. The institutional clinical board approved the study; informed consent was not required. Seven observers independently assessed randomized anonymous digital radiographs of the wrist from 259 subjects; 146 had fractures, and 113 were healthy control subjects (151 male and 108 female subjects; average age, 33 years). Follow-up radiographs and/or computed tomographic scans were used as the reference standard for patients with fractures, and follow-up radiographs and/or clinical history data were used as the reference standard for controls. The PC was a standard hospital machine with a 17-inch (43-cm) color monitor with which Web browser display software was used. The PACS workstation had two portrait 21-inch (53-cm) monochrome monitors that displayed 2300 lines. The observers assigned scores to the radiographs on a scale of 1 (no fracture) to 5 (definite fracture). Receiver operating characteristic (ROC) curves, sensitivity, specificity, and accuracy were compared.

RESULTS: The areas under the ROC curves were almost identical for the PC and workstation (0.910 vs 0.918, respectively; difference, 0.008; 95% confidence interval: –0.029, 0.013). The average sensitivity with the PC was almost identical to that with the workstation (85% vs 84%, respectively), as was the average specificity (82% vs 81%, respectively). The average accuracy (83%) was the same for both.

CONCLUSION: The results of this study showed that there was no difference in accuracy of observer performance for detection of wrist fractures with a PC compared with that with a PACS workstation.

© RSNA, 2005




This article has been cited by other articles:


Home page
Br. J. Radiol.Home page
D S Brettle
Display considerations for hospital-wide viewing of soft copy images
Br. J. Radiol., July 1, 2007; 80(955): 503 - 507.
[Abstract] [Full Text] [PDF]


Home page
RadiologyHome page
B. F. Branstetter IV
Basics of Imaging Informatics: Part 1
Radiology, June 1, 2007; 243(3): 656 - 667.
[Abstract] [Full Text] [PDF]


Home page
Emerg. Med. J.Home page
M H Brem, C Bohner, A Brenning, K Gelse, T Radkow, M Blanke, P M Schlechtweg, G Neumann, I Y Wu, W Bautz, et al.
Evaluation of low-cost computer monitors for the detection of cervical spine injuries in the emergency room: an observer confidence-based study.
Emerg. Med. J., November 1, 2006; 23(11): 850 - 853.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
RADIOLOGY RADIOGRAPHICS RSNA JOURNALS ONLINE
Copyright © 2005 by the Radiological Society of North America.