
View larger version (57K):
[in a new window]
|
Acoustic noise spectra of FGRET and FSPGR at operator location with short repetition time (A,B) and of FGRET at operator location and patient location with long repetition time (TR) (C,D). In A and B, the harmonics are clearly distinguishable, whereas in C and D, the harmonics are too closely spaced, giving rise to a broadband noise spectrum. Note the peak frequency at approximately 2 kHz in D. Spectra were obtained with a 4,096-point fast Fourier transform and a Hannig filter. SPL = sound pressure level.
|