Technical Developments |
1 From the Cardiac Magnetic Resonance Unit (S.P., S.B., T.R.J., G.J.B., M.U.S.) and Department of Medical Physics (J.P.R.), General Infirmary at Leeds, Great George St, Rm 170, D-floor, Jubilee Wing, Leeds LS1 3EX, England. Received February 25, 2002; revision requested April 24; final revision received August 29; accepted September 30. S.P. supported by a British Heart Foundation Research Fellowship. Address correspondence to S.P. (e-mail: sven.plein@leedsth.nhs.uk).
| ABSTRACT |
|---|
|
|
|---|
© RSNA, 2003
Index terms: Heart, MR, 524.12142, 541.12142, 542.12142 Magnetic resonance (MR), functional imaging, 524.12142, 541.12142, 542.12142 Magnetic resonance (MR), motion correction, 524.12144, 541.12144, 542.12144 Magnetic resonance (MR), vascular studies, 524.12142, 541.12142, 542.12142
| INTRODUCTION |
|---|
|
|
|---|
External cuirass-type respirators allow noninvasive regulation of the respiration of nonsedated subjects. They can generate highly reproducible chest wall and diaphragmatic excursions (9,10) and can be triggered by ECG. If the devices are set to perform one respiratory cycle per R-R interval, identical respiratory movements can be generated in every cardiac cycle. When applied to cardiac MR imaging, this may permit a synchronized approach to cardiac and respiratory motion compensation, as data could be acquired in every cardiac cycle without the need for breath holding or respiratory navigator gating. Thus, the efficiency of image acquisition may be improved and imaging times may be shortened.
In this study, we evaluated the feasibility of ECG-synchronized respiration with an external respirator in cardiac MR imaging.
| Materials and Methods |
|---|
|
|
|---|
The Respirator
The external respirator used in this study (RTX; Medivent, London, England) consists of a portable computerized power and control unit, which is attached by a flexible large-bore tube to a cuirass (Fig 1). The cuirass is made of specially formulated and molded clear plastic. It fits over the front of the chest and abdomen to include the umbilicus. The sides of the cuirass are bordered with soft foam rubber to provide a comfortable airtight seal. The standard cuirass was custom-modified for use in the MR environment in this study. The modified cuirass extends mainly anteriorly to contain the cardiac receiver coil and fits closely to the sides of the subject. The power unit generates a cyclic pressure change inside the cuirass, with positive pressure creating inspiration and negative pressure creating expiration. A pressure transducer connected to the inside of the cuirass transmits pressure data to the control unit so that exact pressures and thus diaphragmatic and chest wall positions can be generated and maintained. Both inspiratory and expiratory phases are actively controlled, ensuring that the respiratory positions between repeated cycles are highly reproducible. Pressures up to plus or minus 70 cm H2O and respiratory rates of 11,000 cycles per minute can be generated. For this study, the control unit was programmed so that in the ECG-triggered mode ventilation was synchronized to the R wave. The duration and depth of expiration and inspiration and the delay of the respiratory phases from the R wave could be freely chosen. The device can be set to trigger a respiratory cycle at every cardiac cycle or at other ratios. In this study, the external respirator was used in the ECG-triggered mode for all imaging, with a 1:1 ratio of respiratory and cardiac cycles.
|
|
The exact settings of the respirator varied for each subject and were optimized individually according to heart rate and required depth of respiration. Expiratory pressures ranged from 10 to 15 cm H2O and inspiratory pressures from -10 to -15 cm H2O, as required to achieve sufficient ventilation (ie, to suppress any spontaneous respiration).
The subjects were familiarized with assisted respiration by operating the device outside the magnet bore prior to commencement of imaging.
Acquisition Protocol
MR imaging was performed with a 1.5-T commercially available system (Gyroscan ACS NT; Philips Medical Systems, Best, the Netherlands) equipped with PowerTrak 6000 gradients (23 mT/m, 105 mT/m/msec) and a five-element cardiac phased-array coil.
Representing the two most commonly used respiratory compensation methods, breath-hold multisection multiphase gradient-echo MR images of the left ventricle and respiratory navigatorgated images of the coronary arteries were acquired. Acquisitions were performed in random order both in the conventional fashion while the external respirator was switched off and during synchronized respiration with the device switched on.
All subjects underwent coronary MR angiography of both the left and right coronary arteries. Data acquisition was performed with a three-dimensional segmented k-space gradient-echo sequence, as described by Botnar et al (8) (7/2.1 [repetition time msec/echo time msec]; flip angle, 25°; T2 preparation prepulses; field of view, 400 x 300 mm; matrix, 512 x 384; in-plane spatial resolution, 1.04 x 0.78 mm; 10 contiguous 3-mm-thick sections interpolated during reconstruction to 20 1.5-mm-thick sections). For conventional imaging, data were acquired during free breathing with prospective navigator gating with real-time correction of the three-dimensional volume position in the craniocaudal direction. For respirator-triggered acquisition, data were acquired with the respirator generating one respiratory cycle in every R-R interval. The respiratory trigger delay and expiratory duration were selected to ensure minimal chest wall and diaphragmatic motion during the acquisition phase in mid-diastole (Fig 2). The delay of the respiratory cycles from the R wave at ECG was set to between 80% and 100% of the R-R interval for coronary MR angiography. The expiratory duration was set to 260300 msec (mean, 275 msec), with inspiration set to occur in the remaining R-R interval. During data acquisition, a respiratory navigator was also applied to allow an assessment of the reproducibility of the respiratory movements.
|
To assess subject tolerance of the external respirator, they were asked by one of the investigators (S.P. or S.B.) if the device caused any pain or discomfort either at rest or during assisted respiration. Any such effects were documented. All subjects were also asked if they regarded the device as "tolerable" in the magnet bore.
Image Analysis
The imaging time and (for coronary imaging) the navigator efficiencies were recorded by one of the investigators (S.P. or S.B.). For coronary imaging, the acquisition time excluded the navigator preparation and data reconstruction times. For gradient-echo cine imaging without the respirator operating, time was recorded from the start of the instructions for the first breath hold until completion of the last breath hold. For acquisition with the respirator operating, the total time for acquisition of all sections was measured.
Images were transferred to a stand-alone workstation (EasyVision; Philips Medical Systems) and analyzed independently and in random order by two experienced observers (S.P., S.B.), who were blinded to the acquisition type.
The multiphase cine images were analyzed by consensus of the two observers for overall image quality on a scale of 04: 0, not interpretable; 1, poor (substantial image artifacts or poor definition of the cardiac structures); 2, average (minimal artifacts but all major cardiac structures identifiable); 3, good (no artifacts and all cardiac structures clearly demarcated); and 4, excellent (outstanding delineation of all cardiac structures). One observer (S.P.) then placed standard-sized regions of interest (100 mm2) in both the blood pool and the myocardium. The mean signal intensity (SI) and the SD of the signal intensity in these regions of interest were measured. From these measurements, the signal-to-noise ratios (SNRs) of blood pool and myocardium and the blood-myocardial contrast-to-noise ratio (CNR) were calculated as follows: SNR = Mean SI/SD; CNR = (SIblood - SImyocardium)/[
(SDblood + SDmyocardium)].
For analysis of coronary arteries, image quality was scored by consensus of both observers on a five-point scale: 0, vessel not seen; 1, poor (substantial image artifacts or vessels poorly defined); 2, average (minimal artifacts but vessels easily identifiable); 3, good (no artifacts and vessels clearly demarcated); and 4, excellent (outstanding definition of coronary vessels). The visible length of the combined left main and left anterior descending coronary arteries and of the right coronary arteries was measured by one observer (S.P.) from the outer edge of their respective aortic sinuses on multiplanar reformations. As for the analysis of cine images, one observer (S.P.) placed standard-sized regions of interest (100 mm2) in the blood pool and the myocardium, and the SNR of the blood pool and the CNR between blood pool and myocardium were calculated.
All data are presented as mean plus or minus SD. Formal statistical analysis was not carried out in view of the small sample size of our study population in this initial feasibility study.
| Results |
|---|
|
|
|---|
In three volunteers, considerable artifacts were induced by the moving cuirass superimposed on the T wave at ECG. This was improved sufficiently to allow data acquisition by placing the ECG leads on the backs of the individuals at a position remote from the anterior chest wall motion. However, in two subjects, ECG artifacts remained and caused intermittent loss of synchronization of the respirator with ECG during acquisition of gradient-echo cine images.
Gradient-Echo Cine Imaging
The total acquisition time at gradient-echo cine imaging was shorter with respirator-triggered acquisition than with conventional breath-hold acquisition (Table 1). The mean image quality score was slightly lower for respirator-triggered than for conventional acquisition. The differences in image quality were caused by motion artifacts on the two respirator-triggered data sets, in which there was intermittent misregistration at ECG. These data sets received a lower quality score than that of the corresponding conventional images. SNR and CNR measurements were similar between the two methods (Table 1). Figure 3 shows end-diastolic and end-systolic images acquired during respirator-controlled respiration; they demonstrate comparable image quality. The figure also shows the different chest positions in diastole and systole that were induced by the respirator.
|
|
|
|
| Discussion |
|---|
|
|
|---|
Synchronized Respiration
The external respirator is presently used mainly for the treatment of acute respiratory failure or for respiratory management after cardiac surgery (1825). In this study, the external respirator was used for an integrated approach to respiratory and cardiac motion correction in cardiac MR imaging. With the device, precise and reproducible synchronization of the respiratory cycle to the cardiac cycle was achieved in the MR environment in a group of volunteers. Imaging sequences that normally require navigator-gated or breath-hold techniques were performed with continuous acquisition during synchronized respiration. The external respirator was well tolerated by this group of subjects, who had no prior experience with the device or with MR imaging, and the learning phase with the device was short.
In this small number of subjects, imaging times were lower and imaging efficiencies were higher with synchronized respiration than those with conventional imaging. Time savings were most substantial in navigator-gated acquisitions, in which imaging times were halved because of higher navigator efficiencies. These efficiencies indicated that the respiratory excursions in each cardiac cycle that were induced by the external respirator were highly reproducible. In breath-hold imaging, the respirator technique reduced imaging times by removing the time conventionally taken for breathing between breath holds. Subjective and objective measurements of image quality were comparable between the two acquisition types. This finding suggests that in gradient-echo cine imaging, the chest motion induced by the external respirator does not affect image quality because the motion occurs reproducibly in each cardiac cycle. For coronary MR angiography, we set the external respirator to generate a relatively slow inspiratory phase during data acquisition in diastole. The finding of image quality that was similar to that with conventional technique suggests that the respirator-induced motion during data acquisition was minimal.
Potential applications of cardiac-respiratory motion synchronization in cardiac MR imaging could be substantial. Results in this study seem to indicate that this technique may remove the time constraints of breath holding and overcome the limited efficiency of navigator imaging, which may allow more efficient and faster image acquisition. The technique is independent of the acquisition sequence used; thus, it can be combined with all MR imaging methods available. The technique should be particularly useful in coronary MR angiography and could be used to increase the resolution of coronary artery imaging or to shorten imaging times in long acquisition protocols.
Patients who are unable to hold their breath might benefit from the device. The external respirator allows adjustment of the depth of respiration to each subjects requirements, and patients with dyspnea could receive mild hyperventilation to improve their comfort in the supine position in the MR imager. Continuous monitoring of PO2 and PCO2 is possible to ensure adequate ventilation. Depending on the timing of the respiratory phases relative to the cardiac cycle, chest compression during expiration may also support systolic contraction, and the negative external pressure during inspiration could increase venous return in diastole. These effects will have to be evaluated in future studies.
Limitations
The main problems encountered in this feasibility study were related to ECG artifacts associated with the use of the external respirator in the magnetic field. The respirator causes relatively large chest excursions, and in some subjects, this caused ECG artifacts, which in turn caused misregistration by the respirator. The ECG artifacts could be largely avoided by ensuring minimal movements of the ECG leads during respiration and in some cases by placing them on the subjects back. In two volunteers, however, some artifacts remained even when there was no obvious movement of the ECG leads, which affected the quality of the data acquired. These artifacts are probably related to fast movements of blood or chest wall during respiration with the external respirator. The exact effects of the respirator on blood flow will have to be assessed in future work.
Although this external respirator was generally well tolerated, two volunteers had transient paraesthesia in the hands caused by pressure from the cuirass in the axilla. The cuirass that was used for this initial study has since been optimized further to provide more space in the axilla and anteriorly to accommodate the receiver coil. Despite these modifications, the space taken up by the cuirass will inevitably limit its applicability to larger subjects in the confinement of the bore of an MR imager. The largest subject included in this pilot study weighed 99 kg and had a body mass index of 30.2. It may not be feasible to use the external respirator in the MR environment in subjects who exceed this size.
Another limitation of this pilot work is that only young healthy volunteers were studied. The extent to which our results can be reproduced in patients will have to be determined in future work.
In conclusion, synchronization of respiration to the cardiac cycle with the external respirator in cardiac MR imaging was feasible in this initial study, and the results seem to indicate improved imaging efficiency with similar image quality in comparison with conventional free-breathing and breath-holding techniques. Further investigation of this technique in a larger cohort and in patients is required. Potential applications may include use in routine imaging to reduce imaging times or increase image resolution and in imaging in patients with poor breath-holding ability.
| ACKNOWLEDGMENTS |
|---|
| FOOTNOTES |
|---|
Author contributions: Guarantor of integrity of entire study, S.P.; study concepts and design, S.P., J.P.R., M.U.S.; literature research, S.P.; clinical studies, S.P., S.B., T.R.J.; data acquisition, T.R.J.; data analysis/interpretation, S.P., S.B., G.J.B.; statistical analysis, S.P.; manuscript preparation, S.P., J.P.R., T.R.J., S.B., M.U.S.; manuscript definition of intellectual content and editing, S.P.; manuscript revision/review and final version approval, all authors.