Radiology
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


DOI: 10.1148/radiol.2281020874
This Article
Right arrow Abstract Freely available
Right arrow Figures Only
Right arrow Full Text (PDF)
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Mayo, J. R.
Right arrow Articles by Müller, N. L.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Mayo, J. R.
Right arrow Articles by Müller, N. L.
(Radiology 2003;228:15-21.)
© RSNA, 2003


Special Report

Radiation Exposure at Chest CT: A Statement of the Fleischner Society1

John R. Mayo, MD, John Aldrich, PhD and Nestor L. Müller, MD, PhD

1 From the Department of Radiology, Vancouver General Hospital, 899 W 12th Ave, Vancouver, British Columbia, V5Z 1M9 Canada. Received August 13, 2002; revision requested September 30; revision received November 8; accepted January 6, 2003. Address correspondence to J.R.M. (e-mail: jmayo@vanhosp.bc.ca).


    ABSTRACT
 TOP
 ABSTRACT
 INTRODUCTION
 RADIATION DOSE MEASUREMENT
 CT RADIATION EXPOSURE
 SCANNER RADIATION EFFICIENCY
 USER-SPECIFIED SCAN PARAMETERS
 DOSE REDUCTION IN CHEST...
 CONCLUSION
 REFERENCES
 
The introduction of helical single–detector row computed tomography (CT) and, more recently, multi–detector row CT has greatly increased the clinical indications for CT. Correspondingly, CT examinations now account for greater than one-half of the radiation dose due to medical procedures in the population of North America. The level of CT radiation dose, especially in the pediatric population, is of concern to radiologists, medical physicists, government regulators, and the media. This review addresses this problem with particular reference to radiation dose in chest CT. Specifically it outlines the topics of measurement units used to quantify radiation exposure, factors affecting CT scanner dose efficiency, scanner settings that determine the administered radiation dose, and radiation dose reduction in chest CT. A table of reference dose values is provided. Given the wide variation documented in chest CT radiation exposure, the authors suggest that reference standards be promoted to minimize excessive CT radiation exposure. In addition, further research into the complex relationship between radiation exposure, image quality, and diagnostic accuracy should be encouraged in order to establish the minimum radiation dose necessary to provide adequate diagnostic information for standard clinical questions.

© RSNA, 2003

Index terms: Computed tomography (CT), radiation exposure • Radiations, exposure to patients and personnel • Radiations, measurement • Thorax, CT, 60.1211


    INTRODUCTION
 TOP
 ABSTRACT
 INTRODUCTION
 RADIATION DOSE MEASUREMENT
 CT RADIATION EXPOSURE
 SCANNER RADIATION EFFICIENCY
 USER-SPECIFIED SCAN PARAMETERS
 DOSE REDUCTION IN CHEST...
 CONCLUSION
 REFERENCES
 
The introduction of helical, or spiral, computed tomography (CT) in the late 1980s revolutionized diagnostic medical imaging (14). Single–detector row CT scanners and, more recently, multi–detector row CT scanners markedly increased the clinical indications for CT. As a result there has been a considerable increase in the number of CT examinations performed and in the average scanned volume obtained per examination. Studies in the United States and the United Kingdom have shown an approximately twofold increase in the number of CT examinations performed between the late 1980s and the late 1990s (57). In a study performed at an academic institution in the United States, Mettler et al (5) found that CT accounted for 11.1% of diagnostic radiologic procedures in 1999 compared with 6.1% in 1990. The results of this study showed that CT accounted for about 67% of the total effective radiation dose from diagnostic radiology in 1999. The results of a radiation dose survey performed in the United Kingdom in 1989 showed that CT accounted for 2% of all radiologic examinations and 20% of the effective dose (6). A second survey, in 1995, showed that CT accounted for 4% of all radiologic examinations (7) and 40% of the effective dose, which is a twofold increase. In the study by Mettler et al (5), 11% of the CT examinations were performed in the pediatric population, which was a higher percentage than previously estimated.

The increase in population radiation exposure from CT, particularly in children, has been of concern to radiologists, medical physicists, government regulators, and the media (8). The suggestion that excessive radiation doses are being prescribed for CT has appropriately aroused the attention of the radiologic community (9,10). Radiologists and medical physicists must be attentive to their responsibility to maintain an appropriate balance between diagnostic image quality and radiation dose (11). It has been suggested that in the rapidly evolving field of multi–detector row CT, issues of radiation dose may have been diminished in the quest for increased image quality, diagnostic accuracy, and new imaging techniques (12).

In this communication we outline the determinants of CT radiation dose and discuss the interaction between image quality and radiation dose. Specifically we outline these topics: (a) measurement units used to quantify radiation exposure, (b) parameters that affect CT radiation dose and efficiency, and (c) advances in dose reduction in chest CT. A complete review of radiation dosimetry and bioeffects is beyond the scope of this article, and interested readers are referred to more complete works in these areas (1216).


    RADIATION DOSE MEASUREMENT
 TOP
 ABSTRACT
 INTRODUCTION
 RADIATION DOSE MEASUREMENT
 CT RADIATION EXPOSURE
 SCANNER RADIATION EFFICIENCY
 USER-SPECIFIED SCAN PARAMETERS
 DOSE REDUCTION IN CHEST...
 CONCLUSION
 REFERENCES
 
There are many methods currently in use for quantifying ionizing radiation (Table 1) (17). The fact that several methods exist attests to the complexity of this issue.


View this table:
[in this window]
[in a new window]

 
TABLE 1. Methods of Quantifying Ionizing Radiation

 
The simplest parameter, radiation exposure, is determined by measuring ionization in air caused by the x-ray beam. The measurement unit is coulombs per kilogram (abbreviation, C/kg). It has limited clinical value, as it does not take into account the area irradiated, the penetrating power of the radiation, or the radiation sensitivity of the irradiated organs. From radiation exposure we can calculate the skin entrance dose, which is important when examining deterministic effects such as skin erythema. Although deterministic effects are not encountered in routine CT, they are of potential concern in CT fluoroscopy (12).

Absorbed dose is determined by measuring the energy absorbed per unit mass within an object. The measurement unit is the gray (abbreviation, Gy). Unlike radiation exposure, the gray is dependent on the composition of the object or subject placed in the radiation beam. However, absorbed dose does not account for the differing radiation sensitivity of organs, and it cannot provide a whole-body risk estimate or be used to facilitate comparisons between examinations in different parts of the body. Equivalent dose is a modification of absorbed dose that incorporates weighting factors to account for the different biologic effect of various sources of radiation. For x rays, the radiation weighting factor is 1 and the equivalent dose has the same numerical value as absorbed dose (18).

Effective dose is a measurement that estimates the whole-body dose that would be required to produce the same stochastic risk as the partial-body dose that was actually delivered in a localized radiologic procedure. Effective dose is useful because it allows comparison to other types of radiation exposure such as whole-body radiation exposure secondary to natural background radiation. Effective dose is calculated by summing the absorbed doses to individual organs weighted for their radiation sensitivity (18). The measurement unit is the sievert (abbreviation, Sv). Effective dose has limitations because it represents the radiation detriment for the general population or the specific population of radiation workers and may not be appropriate for many patient populations (19,20). However, it is currently the best measurement available.

Effective dose can be calculated for chest CT by using dose distributions precalculated for specific CT scanner geometry and beam quality (7,2023). These precalculated distributions can be individualized for the CT technical parameters by entering specific tube current, tube voltage, scanned volume, and pitch values. It must be noted that the calculated effective dose values are for reference subjects or phantoms, not for specific patients. Once the effective dose has been calculated, risk estimates for stochastic effects can be produced by using a linear extrapolation of radiation exposure data from Japanese atomic bomb survivors (18,24,25). While the stochastic risk depends on such factors as nationality and age at exposure, the International Commission on Radiological Protection, or ICRP, has recommended the use of a conservative risk of 50 additional fatal cancers induced per million people of the general population exposed to 1 mSv of medical radiation (18). The assessment of stochastic risk is discussed in further detail in ICRP report 60 (18).

Since it is not actually possible to measure the absorbed dose inside a patient, it is not feasible to calculate the exact effective dose for each patient examination. However, it is possible to make a good estimate of the dose and thus the effective dose. In a clinical setting, it would be helpful to be able to estimate the effective dose before the CT examination. This has been made possible recently by the availability on the CT scanners of data derived from measurements made in head and body phantoms. These data are shown on the scanners as the CT dose index (DI) and dose-length product (DLP) and can be used to calculate an effective dose in a reference subject (Fig 1).



View larger version (29K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 1. Diagram shows algorithm for the estimation of radiation exposure risk from CT.

 
The CT DI measurement is made in a plastic phantom designed to simulate either a head (16-cm diameter) or a body (32-cm diameter). To account for imperfect collimation of the x-ray beam and radiation scattered from neighboring scan sections, CT DI measurements are integrated over a 100-mm length. Measurements are made in the center of the phantom and close to the periphery by using an ionization chamber. Beam attenuation causes the dose inside the phantom to change with depth. Therefore, a weighted CT DI is calculated by adding together one-third of the central values and two-thirds of the peripheral values. The DLP depends on the weighted CT DI, the pitch, and the length of the scan. The manufacturer loads tables within the scanner software that provide the weighted CT DI value for the scan parameters specified. The start and end locations of the helical CT study are selected from the localizing image taken before the study and are used to calculate the length of the study. A DLP for that specific CT examination can be calculated by using these values. Table 2 provides general conversion coefficients from DLP to effective dose for typical regions of the body (26). Once the effective dose is calculated, the risk of cancer induction can be estimated by using the population average risk of 50 induced cancers per millisieverts of effective dose per million people exposed (18).


View this table:
[in this window]
[in a new window]

 
TABLE 2. Normalized Effective Dose Coefficients

 
Conversion of effective dose or DLP calculations to radiation risk can be performed for the pediatric population as it can for the adult population (2729). It must be noted that there is a pronounced age effect on the risk of cancer from ionizing radiation (30,31). Children may be an order of magnitude more sensitive than adults to the risk of cancer induction from the same effective dose of ionizing radiation (Fig 2). This arises from the fact that they have more time to express the cancer and the fact that they have more rapidly dividing cells than adults have.



View larger version (27K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 2. Graph compares lifetime mortality risk from cancer per sievert to age at the time of exposure. ICRP = International Commission on Radiological Protection. (Adapted and reprinted, with permission, from reference 70.)

 

    CT RADIATION EXPOSURE
 TOP
 ABSTRACT
 INTRODUCTION
 RADIATION DOSE MEASUREMENT
 CT RADIATION EXPOSURE
 SCANNER RADIATION EFFICIENCY
 USER-SPECIFIED SCAN PARAMETERS
 DOSE REDUCTION IN CHEST...
 CONCLUSION
 REFERENCES
 
The relatively high radiation dose associated with CT results from two properties of the technique (32). First, unlike analogue film radiography, in which the image acquisition and display are both reliant on film, CT is a digital technique in which image acquisition and display can be independently manipulated. Therefore, when CT radiation dose is excessive the image does not become too dark (as it does in film radiography) but instead improves because of decreased image noise (33). Second, visualization of image noise is enhanced by the ability to map the entire gray scale onto selected segments of the CT number scale. As a result, image degradation due to quantum noise (mottle) is easily visible and may interfere with image analysis. As a result of these two effects and the overwhelming drive of radiologists for the very best image quality to support high levels of diagnostic accuracy, CT images are often obtained with high radiation exposure to the patient, which may not be recognized by the radiologist.

In the early 1990s, concern was raised regarding radiation dose in chest CT (3436). Patient radiation-dose surveys showed (6,7,3740) wide variations in radiation exposure between different sites and equipment. These CT data demonstrated that greater consideration needed to be given to optimizing chest CT exposures. However, the current data concerning CT radiation dose indicate that insufficient progress has been made.


    SCANNER RADIATION EFFICIENCY
 TOP
 ABSTRACT
 INTRODUCTION
 RADIATION DOSE MEASUREMENT
 CT RADIATION EXPOSURE
 SCANNER RADIATION EFFICIENCY
 USER-SPECIFIED SCAN PARAMETERS
 DOSE REDUCTION IN CHEST...
 CONCLUSION
 REFERENCES
 
Several physical aspects of CT scanners result in wasted radiation dose. These include the shielding effect of the collimator between the patient and the detector, imperfect collimation of the x-ray beam, and movement of the x-ray focal spot. The sum of all these effects is measured by the geometric efficiency of the CT scanner.

Single–detector row CT scanners, with their wide single–detector row configuration, have higher geometric efficiency than multi–detector row scanners. The decreased geometric efficiency of multi–detector row CT scanners arises from three major factors: the gaps between detector elements in the array, the effect of focal spot penumbra, and the motion of the focal spot. Since the focal spot of the x-ray tube is not a point, the collimator cannot perfectly collimate the beam. Therefore the edge of the beam, or penumbra, has spatially varying x-ray intensity. In single–detector row helical CT scanners this portion of the beam can be detected and used in the reconstruction process. With multi–detector row CT scanners, however, use of the penumbra would result in different readings from detectors in this region compared with those in the central, or umbra, portion of the beam. Therefore the active detectors in multi–detector row CT scanners measure only the umbra of the x-ray beam. Radiation in the penumbra falls on inactive detectors and is discarded, although it contributes to patient radiation dose. In addition, thermal and mechanical stresses within the x-ray tube cause the focal spot to move. As a result, the x-ray beam wanders slightly across the detector array during CT data acquisition. Widening the x-ray beam to compensate for the penumbra and focal spot motion leads to a decrease in the geometric efficiency and an increase in the radiation dose. Because the penumbra is a fixed size, its effect is greatest on four-section CT scanners operating with thin-section collimation. The effect is progressively less severe with eight-, 16-, and 32-section multi–detector row CT scanners. Manufacturers have devised beam-tracking systems to stabilize the position of the x-ray beam and thereby minimize the radiation-wasting effect of focal spot motion (41).

Scattered radiation is formed by the interaction of the primary beam with the body of the patient. Scattered radiation exits the body in all directions, and if detected it reduces contrast and may generate artifacts. In plain chest radiography 50%–90% of film darkening (depending on the technique chosen) is due to scattered radiation, which contributes to the low soft-tissue contrast of this technique (42). The extensive collimation at CT reverses this ratio, and 90% of detected x rays are primary image photons. This partially accounts for the improved soft-tissue contrast at CT.

Because of scatter and imperfect collimation, the radiation intensity profile does not fall to zero at the edge of the nominal section width. It has been shown by using a single-section CT scanner that contiguous sections generate a peak radiation dose approximately 50% greater than that of a single CT section (10-mm collimation, 10-mm table increment, measured at the surface of a 15-cm head phantom) (43). The increase in radiation dose associated with multiple, adjacent CT sections has been measured and is characterized by the multiple-scan average dose, or MSAD, parameter (44,45). Helical CT scanning with a pitch of 1 results in a dose distribution that is essentially equivalent to that of contiguous single–detector row CT imaging (46). Overlapping sections or helical CT scanning with a pitch of less than 1 can result in even higher doses if techniques are not adjusted. Radiation dose can be reduced if gaps are introduced between scanned sections (47,48). However, diagnostic information can be lost by using section gaps because only a portion of the chest is imaged. For this reason, gaps between sections are practical only when diffuse processes such as interstitial lung disease are imaged (49).

CT detectors vary in their efficiency. Ideally a detector should count all incident beam x-ray photons. Depending on the technology used, however, detectors will record only 60% (high-pressure xenon detectors) to 95% (solid-state detectors) of the incident x-ray photons. Most current detectors are solid state. The accuracy of conversion of the absorbed x-ray signal into an electrical signal is known as the conversion efficiency. The overall dose efficiency of the scanner is the product of the geometric efficiency, the quantum detection efficiency, and the conversion efficiency (50). The overall dose efficiency can vary substantially between scanners. Noise is also introduced by the electronics of the data acquisition system of the scanner. The sum of quantum noise and electronic noise results in differences in image quality between scanners at the same radiation dose.

CT is similar to other radiologic techniques in that the primary x-ray beam is filtered to eliminate low-energy photons, which would be preferentially absorbed relative to high-energy photons and contribute to radiation dose. With CT, additional spatially varying filtration is often placed in the primary x-ray beam. These filters reduce (a) the necessary dynamic range of the detector system in the periphery of the detector array and (b) the radiation dose for larger fields of view. They are often referred to as bow tie filters because of their shape, and they create variations in entrance radiation exposure depending on both the size of the object and its position in the field of view. For some CT scanners, multiple filters of varying shapes are moved into place based on the specified field of view of the scan. In other scanners these filters are permanently positioned. These filters substantially reduce radiation dose of CT scans in adult patients, but they are less effective in pediatric patients.


    USER-SPECIFIED SCAN PARAMETERS
 TOP
 ABSTRACT
 INTRODUCTION
 RADIATION DOSE MEASUREMENT
 CT RADIATION EXPOSURE
 SCANNER RADIATION EFFICIENCY
 USER-SPECIFIED SCAN PARAMETERS
 DOSE REDUCTION IN CHEST...
 CONCLUSION
 REFERENCES
 
Reduction in radiation dose results in increased image noise and decreased image quality. Studies assessing the subjective evaluation of chest CT scans have demonstrated that radiologists consistently gave higher image-quality scores to images obtained with a higher radiation dose (51,52). Image noise can be measured by placing a region of interest (>100 pixels) in an area of uniform density in the body (eg, the thoracic aorta) (32). The SD of the pixel values represents image noise. It is noted that the choice of reconstruction algorithm influences image noise, and higher noise is obtained by using high-spatial-frequency reconstruction algorithms (eg, bone or lung algorithms) rather than low-spatial-frequency algorithms (eg, standard, soft-tissue algorithm). High-spatial-frequency reconstruction algorithms are most commonly used when one is searching for fine structures within bones or lung tissue. The increase in image noise associated with the high-spatial-frequency algorithm is not a problem in these applications, because of the high radiographic contrast of these tissues.

Radiation dose and image noise can be modified by adjusting the tube current, scan time, and tube voltage. In practice, the tube current is usually adjusted to change the radiation dose and image noise. In most CT scanners, the tube current is adjustable in steps from 20 mA to approximately 400 mA. The radiation dose can also be linearly affected by scan time, but the time is usually minimized in imaging of the chest to reduce the effect of patient motion. Increasing the tube voltage increases the output of the x-ray tube. If the tube current and scan times are not changed, increasing the tube voltage will increase the radiation dose to the patient. Changes in tube voltage also affect CT tissue attenuation values, which can change tissue contrast in a complex fashion. In practice, tube voltage is not commonly adjusted between patients when chest CT is performed. It is noted that the radiation exposure delivered at a given tube voltage and current setting will vary greatly between CT scanners of different models and manufacturers because of differences in scanner geometry (x-ray tube-to-patient separation) and x-ray tube filtration.

Helical CT scanners introduced a new parameter: pitch. For single–detector row helical CT scanners, pitch is defined as the table travel per 360° x-ray tube rotation divided by the beam collimation (53). In many cases the table feed (eg, 5 mm per x-ray tube rotation) and beam collimation (eg, 5 mm) are identical, and the resultant pitch is 1. This yields one helical turn per section thickness and a radiation exposure equal to that of contiguous transverse sections. However, the table can be made to feed more rapidly (eg, 10 mm per x-ray tube rotation) without changing the beam collimation (5 mm). This results in a pitch of 2. Examinations with pitch values greater than 1 cover larger volumes in shorter times, which provides either reduced motion artifact or thinner sections. Scans obtained with elevated pitch have lower image quality because the section profile is broadened. However, the radiation dose delivered by the examination is decreased by the value of the pitch (eg, one-half of the radiation exposure for a pitch of 2) if the tube voltage and current are kept constant. It should be noted that in many multi–detector row scanners, the tube current is automatically increased to compensate for increased noise at higher pitch values, which may cancel out the radiation dose reduction. In some cases, such as when helical CT is used in the detection of pulmonary embolism, it has been shown that improved image contrast can be obtained with reduced radiation dose by using thinner sections at pitch values of 1.5–2 (3).

One manufacturer of multi–detector row CT scanners has redefined pitch as the table travel divided by the detector aperture (54). This definition elevates the value of the pitch by the number of detector rows. With the standard definition, for example, the acquisition of four 1.25-mm sections at a table speed of 10 mm per second in a 0.5-second scanner results in a pitch of 1. However, by using the same parameters but using the detector aperture as the denominator, the pitch value increases to 4. We agree with others (55) that this new definition of pitch does not demonstrate clearly the relationship between radiation dose and x-ray beam overlap found with the original definition. For this reason we believe that the original definition of pitch is preferable. We also note that the original definition of pitch is being adopted by an international standards committee on CT terminology (56).

In the past, the tube current of CT scanners was uniform at all angles around the patient. However, the chest is an elliptical object that has higher attenuation from left to right than from anterior to posterior. Manufacturers have introduced programs that alter the tube current, which increases radiation dose laterally and decreases it in the thinner anteroposterior direction. This has been shown to decrease radiation dose (5759) with minimal effect on image quality. In the future we believe CT scanners will adjust dose automatically during the scanning process to compensate for the size and density of the body section being scanned, which will result in a signal-to-noise ratio that is adequate for diagnosis but is not excessive.

Repeated scanning of the same region (eg, unenhanced and contrast material enhanced) increases the radiation dose in a linear fashion. Therefore, if unenhanced CT is routinely performed prior to the contrast-enhanced CT, the radiation dose is doubled. This effect can be markedly reduced if the unenhanced CT is a high-resolution study (eg, 1-mm collimation at 10-mm spacing), for which the radiation dose is 10% of that of contiguous conventional CT or helical CT with a pitch of 1.

The tube current and voltage settings are usually set according to local experience and practice. Radiation dose surveys have noted wide variation in these settings between institutions (6,7,3740). To decrease this variation and protect the public from inadvertent overexposure, the European communities have published suggested reference dose values (60) for many CT examinations (Table 3). These reference dose values were obtained by surveying a large number of institutions in Europe and adopting the 75th percentile of responses as the reference dose values. These values serve as a guide to acceptable practice in Europe.


View this table:
[in this window]
[in a new window]

 
TABLE 3. Proposed Reference Dose Values for Routine CT Examinations on the Basis of Absorbed Dose to Air

 

    DOSE REDUCTION IN CHEST CT
 TOP
 ABSTRACT
 INTRODUCTION
 RADIATION DOSE MEASUREMENT
 CT RADIATION EXPOSURE
 SCANNER RADIATION EFFICIENCY
 USER-SPECIFIED SCAN PARAMETERS
 DOSE REDUCTION IN CHEST...
 CONCLUSION
 REFERENCES
 
The concept of reduced tube current for conventional 10-mm collimation chest CT was introduced in 1990 by Naidich et al (61), who demonstrated acceptable image quality for assessment of lung parenchyma with low tube current settings (20 mAs). While these images were adequate for assessing lung parenchyma, they had a considerable increase in noise, which resulted in marked degradation of image quality with mediastinal windows. For this reason the authors noted that such low-dose techniques were most suited for assessment of children and possibly for screening patients at high risk for lung cancer. These recommendations have been implemented and further studied in lung cancer screening programs (6264).

Similar dose reduction strategies have been applied to thin-section CT (also known as high-resolution CT) of the chest, in which no significant difference in lung parenchymal structures was seen between low-dose (40 mAs) and high-dose (400 mAs) thin-section CT images (65). Although differences were not statistically significant, changes in ground-glass opacity were difficult to assess on low-dose images because of the increased image noise. Therefore, it was recommended that 200 mAs should be used for initial thin-section CT and lower doses (ie, 40–100 mAs) should be used for follow-up CT examinations.

The radiation dose associated with thin-section CT has been controversial. DiMarco and Briones (34) quoted the high value of 120–140 mGy, which had been reported in an early article on thin-section CT (66). This dose estimate was measured by using contiguous 1.5-mm sections, 510 mAs, and CT DI methods in a head-sized (16-cm) CT quality control phantom. The CT DI measurement was designed to facilitate dose comparisons between CT scanners and was not appropriate for use in describing the relative dose of thin-section CT. As previously noted, the effective dose is a better measure of radiation dose because it takes into account the significant reduction in radiation risk associated with the noncontiguous sections used in thin-section CT. With 10-mm intersection gaps, the effective dose of thin-section CT is 10% that of either conventional contiguous CT or helical CT with a pitch of 1. The effective dose of thin-section CT is reduced to approximately 5% with the use of 20-mm intersection gaps. As noted previously, low-dose, thin-section CT can also be performed in selected patients. It has been shown that three low-dose, thin-section CT sections provide an effective dose comparable to that of posteroanterior chest radiography (0.05 mSv), with no significant loss of diagnostic accuracy in interstitial lung disease (P > .25) (49).

The relationship between radiation exposure and image quality with both mediastinal and lung windows has been evaluated on conventional 10-mm collimation chest CT images (51) on a single model of CT scanner. Although findings of this study showed a consistent increase in mean image quality with higher radiation exposure, they did not show a significant difference in the detection of mediastinal or lung parenchymal abnormalities from 20 to 400 mAs. The authors concluded that with the CT scanner model they used, adequate image quality could be consistently obtained in average-sized patients by using tube currents of 100–200 mAs. This study was limited by the small number of patients (n = 30), the specific CT scanner factors (geometry, filtration, tube voltage), and the experimental design, which limited low-dose sections to two levels that often were not those with clinically relevant findings. The authors noted that to evaluate further the effect of reduced radiation dose on diagnostic accuracy in chest CT, comparison of complete chest CT studies at a variety of radiation exposures in a large number of patients would be required. However, they noted that such a study could not be performed in patients because of the unacceptable radiation dose that would result from multiple CT examinations at differing radiation exposures. Additionally, the variable effect of motion artifacts on repeated scanning would make comparison difficult.

A practical method for evaluating the effect of reduced radiation dose on image quality is computer simulation (67). The technique consists of obtaining a diagnostic scan with standard dose and then modifying the raw scan data by adding Gaussian-distributed random noise to simulate the increased noise associated with reduced radiation exposure. The raw scan data are then reconstructed by using the same field of view and reconstruction algorithm as the high-dose reference scan. In a validation trial, experienced chest radiologists were unable to distinguish simulated reduced-dose CT images from real reduced-dose CT images (67). Computer simulation allows investigators to determine the effect of dose reduction on diagnostic accuracy without exposing patients to radiation unnecessarily. In addition, the simulated images are in exact registration with the original images, eliminating artifacts due to volume averaging or motion. This technique has been used recently to evaluate the diagnostic effect of radiation dose reduction in pediatric abdominal CT (68).

Finally, it should be noted that although CT is a modality with relatively high radiation dose, in some cases CT has replaced modalities with higher radiation exposures such as pulmonary angiography or bronchography (Table 4).


View this table:
[in this window]
[in a new window]

 
TABLE 4. Comparison of Effective Doses

 

    CONCLUSION
 TOP
 ABSTRACT
 INTRODUCTION
 RADIATION DOSE MEASUREMENT
 CT RADIATION EXPOSURE
 SCANNER RADIATION EFFICIENCY
 USER-SPECIFIED SCAN PARAMETERS
 DOSE REDUCTION IN CHEST...
 CONCLUSION
 REFERENCES
 
The introduction of helical and multi–detector row CT scanners has resulted in an increase in the number of indications for and diagnostic accuracy of CT. However, the current level of radiation exposure from CT is high, particularly in the pediatric population. Radiation dose surveys have indicated that there is large variation in the technical factors employed by physicians, and there is a resultant large variation in the radiation dose to patients. Reference dose values for chest CT have been developed and published. Radiologic societies should be encouraged to consider the benefits that adoption of such standards would have on medical practice. Further research into the complex relationship between radiation exposure, image noise, and diagnostic accuracy should be encouraged to establish scientifically the minimum radiation doses that provide adequate diagnostic information for standard clinical questions. Once these minimum levels of image quality are determined and validated, automatic exposure controls for CT scanners should be developed to ensure that all patients undergo CT with techniques that conform to the ALARA (As Low As Reasonably Achievable) principle. Radiologists must take the lead in promoting these measures for patient protection. Since children have an increased radiation sensitivity, initial dose reduction efforts should be focused on the pediatric population. The Society for Pediatric Radiology has recently made recommendations in this regard (7385). Finally, it is noted that the complexity of CT requires a close collaboration between radiologists and medical physicists to successfully reduce radiation dose while maintaining diagnostic accuracy.


    FOOTNOTES
 
Abbreviations: DI = dose index, DLP = dose-length product


    REFERENCES
 TOP
 ABSTRACT
 INTRODUCTION
 RADIATION DOSE MEASUREMENT
 CT RADIATION EXPOSURE
 SCANNER RADIATION EFFICIENCY
 USER-SPECIFIED SCAN PARAMETERS
 DOSE REDUCTION IN CHEST...
 CONCLUSION
 REFERENCES
 

  1. Kalender WA, Seissler W, Klotz E, Vock P. Spiral volumetric CT with single-breath-hold technique, continuous transport, and continuous scanner rotation. Radiology 1990; 176:181-183.[Abstract/Free Full Text]
  2. Kalender WA, Polacin A. Physical performance characteristics of spiral CT scanning. Med Phys 1991; 18:910-915.[CrossRef][Medline]
  3. Remy-Jardin M, Remy J, Mayo JR, Müller NL. Acquisition, injection, and reconstruction techniques In: CT angiography of the chest. Philadelphia, Pa: Lippincott Williams & Wilkins, 2001; 1-14.
  4. Berland LL, Smith JK. Multidetector-array CT: once again, technology creates new opportunities. Radiology 1998; 209:327-329.[Free Full Text]
  5. Mettler FA, Jr, Wiest PW, Locken JA, Kelsey CA. CT scanning: patterns of use and dose. J Radiol Prot 2000; 20:353-359.[CrossRef][Medline]
  6. Shrimpton PC, Jones DG, Hillier MC, Wall BF, Le Heron JC, Faulkner K. Survey of CT practice in the UK II. Dosimetric aspects. NRPB report no. R-249. Chilton, England: National Radiological Protection Board, 1991.
  7. Shrimpton PC, Edyvean S. CT scanner dosimetry. Br J Radiol 1998; 71:1-3.[Medline]
  8. Sternberg S. CT scans in children linked to cancer later. USA Today 2001; Jan 22:1.
  9. Rogers L. Radiation exposure in CT: why so high? AJR Am J Roentgenol 2001; 177:277.[Free Full Text]
  10. Rogers LF. Serious business: radiation safety and radiation protection. AJR Am J Roentgenol 2001; 177:1.[Free Full Text]
  11. Nickoloff EL, Alderson PO. Radiation exposures to patients from CT: reality, public perception and policy. AJR Am J Roentgenol 2001; 177:285-287.[Free Full Text]
  12. Golding SJ, Shrimpton PC. Radiation dose in CT: are we meeting the challenge?(commentary). Br J Radiol 2002; 75:1-4.[Free Full Text]
  13. Mettler FA, Jr, Upton AC. Medical effects of ionizing radiation 2nd ed. Philadelphia, Pa: Saunders, 1995.
  14. Huda W, Atherton JV. Energy imparted in computed tomography. Med Phys 1995; 22:1263-1269.[CrossRef][Medline]
  15. Jones AP, Mott DJ, Parkinson L. Experience with a new simple method for the determination of doses in computed tomography. Radiat Prot Dosimetry 1992; 43:139-142.[Abstract]
  16. Metz CE, Wagner RF, Doi K, Brown DG, Nishikawa RM, Myers KJ. Toward consensus on quantitative assessment of medical imaging systems. Med Phys 1995; 22:1057-1061.[CrossRef][Medline]
  17. Huda W. Radiation dosimetry in diagnostic radiology. AJR Am J Roentgenol 1997; 169:1487-1488.[Free Full Text]
  18. Recommendations of the International Commission on Radiological Protection. ICRP publication no 60. Oxford, England: Pergamon, 1991.
  19. Sinclair WK. The concept of effective dose. Health Phys 1994; 66:586.[Medline]
  20. McCollough CH, Schueler BA. Calculation of effective dose. Med Phys 2000; 27:828-837.[CrossRef][Medline]
  21. Jones DG, Shrimpton PC. Normalised organ doses for x-ray CT calculated using Monte Carlo techniques NRPB-SR250. Chilton, England: National Radiological Protection Board, 1993.
  22. Atherton JV, Huda W. Energy imparted and effective dose in computed tomography. Med Phys 1996; 23:735-741.[CrossRef][Medline]
  23. Zankl M, Panzer W, Drexler G. The calculation of dose from external photon exposures using reference human phantoms and Monte Carlo methods. VI. Organ doses from computed tomographic examinations Neuherberg, Germany: GSF-Forschungszentrum für Umwelt und Gesundtheit, Institut für Strahlenschutz, 1991.
  24. Beir V. Health effects of exposure to low levels of ionizing radiation Washington, DC: National Academy Press, 1990.
  25. Risk estimates for radiation protection. NCRP report no 115. Bethesda, Md: National Council on Radiation Protection and Measurements, 1993.
  26. Jessen KA, Shrimpton PC, Geleijns J, Panzer W, Tosi G. Dosimetry for optimisation of patient protection in computed tomography. Appl Radiat Isot 1999; 50:165-172.[CrossRef][Medline]
  27. Chapple C-L, Willis S, Frame J. Effective dose in paediatric computed tomography. Phys Med Biol 2002; 47:107-115.[CrossRef][Medline]
  28. Shrimpton PC, Wall BF. Reference doses for paediatric computed tomography. Radiat Prot Dosim 2000; 90:249-252.[Abstract]
  29. Zankl M, Panzer W, Drexler G. Tomographic anthropomorphic models. II. Organ doses from computed tomographic examinations in paediatric radiology GSF-Forschungszentrum für Umwelt und Gesundtheit, Institut für Strahlenschutz: Neuherberg, Germany, 1993.
  30. Pierce DA, Shimizu Y, Preston DL, Vaeth M, Mabuchi K. Studies of the mortality of atomic bomb survivors. Report 12, part I. Cancer: 1950–1990. Radiat Res 1996; 146:1-27.[Medline]
  31. Brenner DJ, Elliston CD, Hall EJ, Berdon WE. Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR Am J Roentgenol 2001; 176:289-296.[Abstract/Free Full Text]
  32. Sprawls P, Jr. AAPM tutorial: CT image detail and noise. RadioGraphics 1992; 12:1041-1046.[Abstract]
  33. Rothenberg LN, Pentlow KS. Radiation dose in CT. RadioGraphics 1992; 12:1225-1243.[Abstract]
  34. Di Marco AF, Briones B. Is chest CT performed too often? Chest 1993; 103:985-986.
  35. Naidich DP, Pizzarello D, Garay SM, Müller NL. Is thoracic CT performed often enough? Chest 1994; 106:331-332.[Free Full Text]
  36. DiMarco AF, Renston JP. In search of the appropriate use of chest computed tomography. Chest 1994; 106:332-333.[Free Full Text]
  37. Nishizawa K, Maruyama T, Takayama M, Okada M, Hachiya J, Furuya Y. Determinations of organ doses and effective dose equivalents from computed tomographic examination. Br J Radiol 1991; 64:20-28.[Abstract]
  38. Conway BJ, McCrohan JL, Antonsen RG, Rueter FG, Slayton RJ. Average radiation dose in standard CT examinations of the head: results of the 1990 NEXT study. Radiology 1992; 184:135-140.[Abstract/Free Full Text]
  39. Shrimpton PC, Hillier MC, Wall BF, et al. Survey of CT practice in the UK I. Aspects of examination frequency and quality assurance. NRPB report no. R-248. Chilton, England: National Radiological Protection Board, 1991.
  40. Panzer W, Scheurer C, Zankl M. Dose to patients in computed tomography examinations: results and consequences from a field study in the Federal Republic of Germany In: Moores BM, Wall BF, Eriskat H, Schibilla H, eds. Optimization of image quality and patient exposure in diagnostic radiology. Report no. 20. London, England: British Institute of Radiology, 1989.
  41. Toth TL, Bromberg NB, Pan TS, et al. A dose reduction x-ray beam positioning system for high-speed multislice CT scanners. Med Phys 2000; 27:2659-2668.[CrossRef][Medline]
  42. Curry TS, III, Dowdey JE, Murray RC, Jr. Christensen’s introduction to the physics of diagnostic radiology 3rd ed. Philadelphia, Pa: Lea & Febiger, 1984; 73.
  43. Pentlow KS, Beattie JW, Laughlin JS. Parameters and design considerations for tomographic transmission scanners. In: Ter-Pogossian MM, Phelps ME, eds. Reconstructive tomography in diagnostic radiology and nuclear medicine. Baltimore, Md: University Park Press, 1977; 267-279.
  44. Shope TB, Gagne RM, Johnson GC. A method for describing the doses delivered by transmission x-ray computed tomography. Med Phys 1981; 8:488-495.[CrossRef][Medline]
  45. Jucius RA, Kambic GX. Radiation dosimetry in computed tomography. Appl Opt Instrum Eng Med 1977; 127:286-295.
  46. McGee PL, Humphreys S. Radiation dose associated with spiral computed tomography. Can Assoc Radiol J 1994; 45:124-129.[Medline]
  47. Evans SH, Davis R, Cooke J, Anderson W. A comparison of radiation doses to the breast in computed tomographic chest examinations for two scanning protocols. Clin Radiol 1989; 40:45-46.[CrossRef][Medline]
  48. Mayo JR, Jackson SA, Müller NL. High-resolution CT of the chest: radiation dose. AJR Am J Roentgenol 1993; 160:479-481.[Abstract/Free Full Text]
  49. Lee KS, Primack SL, Staples CA, Mayo JR, Aldrich JE, Müller NL. Chronic infiltrative lung disease: comparison of diagnostic accuracies of radiography and low- and conventional-dose thin-section CT. Radiology 1994; 191:669-673.[Abstract/Free Full Text]
  50. Cunningham IA. Computed tomography: instrumentation. In: Bronzino JE, eds. The biomedical engineering handbook. Boca Raton, Fla: CRC Press, 1995; 990-1002.
  51. Mayo JR, Hartman TE, Lee KS, Primack SL, Vedal S, Müller NL. CT of the chest: minimal tube current required for good image quality with the least radiation dose. AJR Am J Roentgenol 1995; 164:603-607.[Abstract/Free Full Text]
  52. Haaga JR, Miraldi F, MacIntyre W, LiPuma JP, Bryan PJ, Wiesen E. The effect of mAs variation upon computed tomography image quality as evaluated by in vivo and in vitro studies. Radiology 1981; 138:449-454.[Abstract/Free Full Text]
  53. Polacin A, Kalender WA, Marchal G. Evaluation of section sensitivity profiles and image noise in spiral CT. Radiology 1992; 185:29-35.[Abstract/Free Full Text]
  54. Hu H. Multi-slice helical CT: scan and reconstruction. Med Phys 1999; 26:1-18.[CrossRef][Medline]
  55. McCollough CH, Zink FE. Performance evaluation of a multi-slice CT system. Med Phys 1999; 26:2223-2230.[CrossRef][Medline]
  56. International Electrotechnical Commission. Amendment 1 to IEC 60601–2–44, edition 2: medical electrical equipment, part 2–44—particular requirements for the safety of x-ray equipment for computed tomography Final draft International Standard. Geneva, Switzerland: International Electrotechnical Commission, 2002; 1-7.
  57. Kalender WA, Wolf H, Suess C, Gies M, Greess H, Bautz WA. Dose reduction in CT by on-line tube current control: principles and validation on phantoms and cadavers. Eur Radiol 1999; 9:323-328.[CrossRef][Medline]
  58. Kalender WA, Wolf H, Suess C. Dose reduction in CT by anatomically adapted tube current modulation. II. Phantom measurements. Med Phys 1999; 26:2248-2253.
  59. Greess H, Wolf H, Baum U, et al. Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions. Eur Radiol 2000; 10:391-394.[CrossRef][Medline]
  60. European guidelines on quality criteria for computer tomography. EUR 16262EN Luxembourg: Office for Official Publication of the European Communities, 2000.
  61. Naidich DP, Marshall CH, Gribbin C, Arams RS, McCauley DI. Low-dose CT of the lungs: preliminary observations. Radiology 1990; 175:729-731.[Abstract/Free Full Text]
  62. Henschke CI, McCauley DI, Yankelevitz DF, et al. Early Lung Cancer Action Project: overall design and findings from baseline screening. Lancet 1999; 354:99-105.[CrossRef][Medline]
  63. Itoh H, Ikeda M, Arahata S, et al. Lung cancer screening: minimum tube current required for helical CT. Radiology 2000; 215:175-183.[Abstract/Free Full Text]
  64. Swensen SJ, Jett JR, Sloan JA, et al. Screening for lung cancer with low-dose spiral computed tomography. Am J Respir Crit Care Med 2002; 165:508-513.[Abstract/Free Full Text]
  65. Zwirewich CV, Mayo JR, Müller NL. Low-dose high-resolution CT of lung parenchyma. Radiology 1991; 180:413-417.[Abstract/Free Full Text]
  66. Mayo JR, Webb WR, Gould R, et al. High-resolution CT of the lungs: an optimal approach. Radiology 1987; 163:507-510.[Abstract/Free Full Text]
  67. Mayo JR, Whittall KP, Leung AN, et al. Simulated dose reduction in conventional chest CT: validation study. Radiology 1997; 202:453-457.[Abstract/Free Full Text]
  68. Frush DP, Slack CC, Hollingsworth CL, et al. Computer-simulated radiation dose reduction for abdominal multidetector CT of pediatric patients. AJR Am J Roentgenol 2002; 179:1107-1113.[Abstract/Free Full Text]
  69. Mayo JR, Aldrich JE. Radiation exposure. In: Remy-Jardin M, Remy J, eds. Spiral CT of the chest. Berlin, Germany: Springer-Verlag, 1996; 39.
  70. United Nations Scientific Committee on the Effects of Atomic Radiation. Sources and effects of ionizing radiation: UNSCEAR 1993 report to the General Assembly, with scientific annexes New York, NY: United Nations, 1993.
  71. Documents of the NRPB National Radiological Protection Board. III Chilton, England: National Radiological Protection Board, 1992; 1-16.
  72. Hart D, Jones DG, Wall DF. Estimation of effective dose in diagnostic radiology from entrance surface dose and dose–area product measurements NRPB report no. R-262. Chilton, England: National Radiological Protection Board, 1994.
  73. Slovis TL. The ALARA concept in pediatric CT: myth or reality? Radiology 2002; 223:5-6.[Free Full Text]
  74. Slovis TL. Welcome and introduction to the ALARA conference on the ALARA concept in pediatric CT intelligent dose reduction. Pediatr Radiol 2002; 32:222.[CrossRef]
  75. Brenner DJ. Estimating cancer risks from pediatric CT: going from the qualitative to the quantitative. Pediatr Radiol 2002; 32:228-231.[CrossRef][Medline]
  76. Ron E. Ionizing radiation and cancer risk: evidence from epidemiology. Pediatr Radiol 2002; 32:232-237.[CrossRef][Medline]
  77. Fearon T. CT dose parameters and their limitations. Pediatr Radiol 2002; 32:246-249.[CrossRef][Medline]
  78. Nickoloff E. Current adult and pediatric CT doses. Pediatr Radiol 2002; 32:250-260.[CrossRef][Medline]
  79. Toth TL. Dose reduction opportunities for CT scanners. Pediatr Radiol 2002; 32:261-267.[CrossRef][Medline]
  80. Brody AS. CT scanner design and patient radiation exposure. Pediatr Radiol 2002; 32:268-271.[CrossRef][Medline]
  81. Huda W. Effective doses to adult and pediatric patients. Pediatr Radiol 2002; 32:272-279.[CrossRef][Medline]
  82. Varchena V. Pediatric phantoms. Pediatr Radiol 2002; 32:280-284.[CrossRef][Medline]
  83. Donnelly LF. Lessons from history. Pediatr Radiol 2002; 32:287-292.[CrossRef][Medline]
  84. Frush DP. Strategies of dose reduction. Pediatr Radiol 2002; 32:293-297.[CrossRef][Medline]
  85. Slovis TL. Discussion, consensus, and mandate. Pediatr Radiol 2002; 32:301-313.[CrossRef][Medline]



This article has been cited by other articles:


Home page
RadiologyHome page
J. Vikgren, S. Zachrisson, A. Svalkvist, A. A. Johnsson, M. Boijsen, A. Flinck, S. Kheddache, and M. Bath
Comparison of Chest Tomosynthesis and Chest Radiography for Detection of Pulmonary Nodules: Human Observer Study of Clinical Cases
Radiology, December 1, 2008; 249(3): 1034 - 1041.
[Abstract] [Full Text] [PDF]


Home page
RadiologyHome page
M.-P. Revel, J.-B. Faivre, T. Letourneau, H. Henon, D. Leys, V. Delannoy-Deken, M. Remy-Jardin, and J. Remy
Patent Foramen Ovale: Detection with Nongated Multidetector CT
Radiology, October 1, 2008; 249(1): 338 - 345.
[Abstract] [Full Text] [PDF]


Home page
ThoraxHome page
A U Wells, N Hirani, and on behalf of the BTS Interstitial Lung Disease Gui
Interstitial lung disease guideline
Thorax, September 1, 2008; 63(Suppl_5): v1 - v58.
[Full Text] [PDF]


Home page
Am. J. Roentgenol.Home page
C. Herzog, D. M. Mulvihill, S. A. Nguyen, G. Savino, B. Schmidt, P. Costello, T. J. Vogl, and U. J. Schoepf
Pediatric Cardiovascular CT Angiography: Radiation Dose Reduction Using Automatic Anatomic Tube Current Modulation
Am. J. Roentgenol., May 1, 2008; 190(5): 1232 - 1240.
[Abstract] [Full Text] [PDF]


Home page
RadiologyHome page
A. B. Smith, W. P. Dillon, B. C. Lau, R. Gould, F. R. Verdun, E. B. Lopez, and M. Wintermark
Radiation Dose Reduction Strategy for CT Protocols: Successful Implementation in Neuroradiology Section
Radiology, May 1, 2008; 247(2): 499 - 506.
[Abstract] [Full Text] [PDF]


Home page
Eur Respir JHome page
M. Cazzola, W. MacNee, F. J. Martinez, K. F. Rabe, L. G. Franciosi, P. J. Barnes, V. Brusasco, P. S. Burge, P. M. A. Calverley, B. R. Celli, et al.
Outcomes for COPD pharmacological trials: from lung function to biomarkers
Eur. Respir. J., February 1, 2008; 31(2): 416 - 469.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Roentgenol.Home page
T. Kubo, P.-J. P. Lin, W. Stiller, M. Takahashi, H.-U. Kauczor, Y. Ohno, and H. Hatabu
Radiation Dose Reduction in Chest CT: A Review
Am. J. Roentgenol., February 1, 2008; 190(2): 335 - 343.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Roentgenol.Home page
J. D. MacKenzie, J. Nazario-Larrieu, T. Cai, M. S. Ledbetter, M. A. Duran-Mendicuti, P. F. Judy, and F. J. Rybicki
Reduced-Dose CT: Effect on Reader Evaluation in Detection of Pulmonary Embolism
Am. J. Roentgenol., December 1, 2007; 189(6): 1371 - 1379.
[Abstract] [Full Text] [PDF]


Home page
RadiologyHome page
M. Remy-Jardin, M. Pistolesi, L. R. Goodman, W. B. Gefter, A. Gottschalk, J. R. Mayo, and H. D. Sostman
Management of Suspected Acute Pulmonary Embolism in the Era of CT Angiography: A Statement from the Fleischner Society
Radiology, November 1, 2007; 245(2): 315 - 329.
[Full Text] [PDF]


Home page
Am. J. Neuroradiol.Home page
A.B. Smith, W.P