|
|
||||||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Thoracic Imaging |
1 From the Department of Radiology, Gazi University School of Medicine, Kat, Besevler, Ankara, Turkey. Received May 29, 2002; revision requested July 30; final revision received January 31, 2003; accepted May 14. Address correspondence to M.A. (e-mail: meharac@yahoo.com).
| ABSTRACT |
|---|
|
|
|---|
MATERIALS AND METHODS: Fifty patients were enrolled in the study. All images were obtained with a scanner capable of 1-second revolution time. Routine lung thin-section CT examination was performed with images reconstructed with bone algorithm. Multiple-segment images reconstructed with lung algorithm were obtained for three levels in the left paracardiac region. Segment images were reconstructed retrospectively with data for 225° rotation rather than the 360° rotation used for a complete scan. To minimize differences resulting from reconstruction algorithms, additional nonsegmented reconstruction was performed with lung algorithm. Three radiologists reviewed each set of images and assigned a quality score. Multiway analysis of variance was performed to compare motion artifact reduction with 225° and 360° reconstructions.
RESULTS: Differences were not significant (P > .05) between scores for images reconstructed with bone or lung algorithms. Differences were significant between scores for reconstructed images obtained with the combination of 360° bone and 225° segment algorithms (P < .001) and for those obtained with the combination of 360° lung and 225° segment algorithms (P < .001).
CONCLUSION: Multiple-segment reconstruction of lung thin-section CT images is an effective technique for reducing cardiac-motion artifacts without increasing patient dose.
© RSNA, 2003
Index terms: Computed tomography (CT), artifact Computed tomography (CT), image processing, 60.12115, 60.12118 Lung, CT, 60.12115, 60.12118
| INTRODUCTION |
|---|
|
|
|---|
Technical developments in the past decade have helped decrease the scan time for helical CT scanners. With most of the currently available scanners, a gantry revolution can be performed in 1 second or less. Despite the reduced acquisition time, however, the quality of thin-section CT images is often compromised because of cardiac-motion artifacts, especially when small anatomic structures (ie, lung vessels, bronchi) or small abnormal structures are imaged next to the heart. Cardiac-motion artifacts may result in double images of pulmonary vessels and fissures and can mimic the appearance of bronchiectasis (2).
Reduced acquisition time (<1 second) and cardiac gating are two methods that can be used alone or in combination to reduce cardiac-motion artifacts, but they are not widely available. A third technique is the use of multiple-segment reconstruction options, which are available on most currently available scanners. The latter technique is a retrospective reconstruction method that allows segmentation of scan data into time-related images and reconstruction of multiple images from a single scan acquisition. Diagrams in Figure 1 show the relationship between multiple-segment reconstruction and cardiac cycle. The purpose of our prospective study was to evaluate the effect of multiple-segment reconstruction on the reduction of cardiac-motion artifacts on lung thin-section CT images.
|
|
| MATERIALS AND METHODS |
|---|
|
|
|---|
|
Segment images were reconstructed with data from only a 225° rotation, rather than the 360° rotation used for a complete 1-second scan. Therefore, the reconstructed images contained data from 0.6 second of the 1-second scan time. For each level, six-segment reconstructed images were obtained. For the same level, each segment image was reconstructed with a delay of approximately 166 msec. The use of other reconstruction options, such as half scanning, was also considered. Because of the fan beam angle in this option, the same 225° data were used to reconstruct a single image compared with the six images obtained with multiple-segment reconstruction. This decrease in the total number of reconstructed images could dramatically alter the capability of reducing cardiac-motion artifacts. Therefore, no other technique was used in this study.
Of the six images obtained with multiple-segment reconstruction, the one with the lowest degree of cardiac-motion artifact was chosen. The criteria for choosing one segment image from among the other segment images for the same level was the clarity of the lung parenchyma, fissure, cardiac contours, small vessels, and bronchi. For this purpose, all six-segment images were reviewed by an experienced radiologist (S.A.), and the image with the lowest degree of motion artifact was selected. This radiologist did not participate in image scoring.
Neither highspatial-frequency bone nor sharp reconstruction algorithms were available with segment reconstruction. Segment images were available with different algorithms, such as the lung kernel. To minimize differences as a result of the reconstruction algorithms and to allow a better comparison, an additional nonsegmented (360°-rotation) reconstruction was obtained with lung algorithm. As a result, sets of images were available for each level: Each set consisted of one 360°-rotation thin-section CT image reconstructed with bone algorithm, a second 360°-rotation thin-section CT image reconstructed with lung algorithm, and a third 225°-rotation segment image chosen from the six-segment images reconstructed with lung algorithm. All sets of images for each level were acquired in random order with lung windows (level, -600 HU; width, 1,500 HU). The time for multiple-segment reconstruction and image acquisition was measured.
Image Analysis
Three radiologists (M.A., A.Y.O., H.C.), who were blinded to the reconstruction technique, independently reviewed each set of images and assigned a quality score to each. Before they scored the images, the readers observed two scans that were obtained in patients who were not included in the study, and they assigned a score together to ensure consistency in use of the grading scheme.
Quality scores were assessed on a four-point scale from 1 (no motion artifact) to 4 (excessive motion artifacts). Clarity of the lung parenchyma, fissure, cardiac contours, small vessels, and bronchi formed the basis for the judgement of motion. Because only patients with good ability to hold their breath were included in the study, motion artifacts were considered to be of cardiac origin.
Statistical Analysis
Scores attributed by each reader for different reconstruction techniques were averaged and expressed as the mean plus or minus SD. Statistical analysis was performed (SPSS, version 10.0; SPSS, Chicago, Ill). Multiway analysis of variance was performed between reconstruction techniques to evaluate reduction in motion artifacts, and
scores were calculated to assess interreader agreement. A P value of less than .05 was considered to indicate a statistically significant difference.
| RESULTS |
|---|
|
|
|---|
scores ranged between 0.32 and 0.56 (P < .05), which reflects a moderate to good interobserver agreement. A set of representative CT images for each technique is presented in Figures 2 and 3.
|
|
|
|
|
|
| DISCUSSION |
|---|
|
|
|---|
Thin-section CT is now accepted as the method of choice for evaluation of lung parenchyma. Temporal resolution has improved with widespread use of spiral scanners capable of 1-second scan time (5). Despite the reduced acquisition time, however, the quality of thin-section CT images is often blurred by cardiac-motion artifacts when small structures next to the heart are imaged. These artifacts may result in double images of small bronchi, pulmonary vessels, and fissures, as well as doubling or blurring of the cardiac border, low-attenuating areas adjacent to the vessel, star artifacts, dark spots, and distortion of normal structures. These motion artifacts degrade image quality and can sometimes mimic bronchiectasis (2).
Corrective reconstruction, postprocessing, tube alignment, fast scanning, and electrocardiographic gating are methods used to reduce motion artifacts (69). Ritchie et al (4) stated that scan times of 0.6 second or electron-beam CT with a scan time of 100 msec are not sufficient to eliminate all motion artifacts. Complete elimination of motion artifacts necessitates an acquisition time of less than 19.1 msec (4). This scan time is far from the capability of currently available CT scanners, even the newest multisection machines.
In the present study, multiple-segment reconstructed images were obtained with data from a 225° rotation instead of the 360° rotation for a complete 1-second scan. In this way, temporally overlapping images were obtained at 166-msec intervals, with each image containing data from the 0.6-second period. The success of motion-artifact reduction increased when cardiac systole occupied a small part in the data used for image reconstruction. Cardiac systole typically occupies approximately 30% of the cardiac cycle (10). With an average heart rate of 83 beats per minute, the systolic phase occupies approximately 220 msec. In the best case scenario in the present study, at least one of the multiple-segment reconstructed images was free of the systolic phase when data collection began just after cardiac systole was completed. On the other hand, in the worst case scenario in the current study, images obtained with a 1-second scan time were affected by cardiac-motion artifacts when data collection began simultaneously with systole. Even in the worst case, however, one of the six-segment reconstructed images contained a smaller part of systole than was depicted on a 1-second scan, which contained two complete systoles. This explains the success in reduction of cardiac-motion artifact on segment reconstructed images.
Bone or other highspatial-frequency algorithms were not available with segment reconstruction. To prevent faulty evaluation of motion artifacts as a result of different image reconstruction algorithms, nonsegmented (360°) images reconstructed with lung algorithm were also included in the comparison for each level. Differences in terms of motion artifacts between thin-section CT images reconstructed with bone algorithm and nonsegmented (360°) images reconstructed with lung algorithm were not significant. When segment images reconstructed with lung algorithm were compared with images obtained with each of the other two techniques, a statistically significant reduction in motion artifacts was found (P < .001).
An important issue with multiple-segment reconstructed images is noise. Increased noise can cause diagnostic problems in the detection of ground-glass areas. Nevertheless, findings in several studies show that with increased noise, there is only minimal compromise in diagnostic information, which is not likely to affect image quality in the lung (11). Additionally, with the multiple-segment reconstruction technique for thin-section CT scans used in the current study, images obtained with bone algorithm and those obtained with multiple-segment technique are obtained simultaneously. If there is any diagnostic ambiguity, the 360° thin-section CT images can be used.
An advantage of segment reconstruction is that no additional scanning is required, and there is no increase in patient dose. The whole procedure is completed without any increase in cost or scan time. The only cost is the additional time for postprocessing. A trained technician can complete all the postprocessing and acquisition of the multiple-segment reconstructed images in 2 minutes. With these benefits, segment reconstructed images can be used successfully to reduce cardiac-motion artifacts in thin-section CT images of the lung. For every thin-section CT image, six segment reconstructions are provided. This can result in an increased number of images stored for each patient. Therefore, dynamic evaluation of thin-section CT images and selection of those that contain motion artifacts suitable for multiple-segment reconstruction is advisable for clinics with a high workload.
Findings in this study demonstrate that multiple-segment reconstructions of thin-section CT images show decreased cardiac-motion artifacts in lung parenchyma. This option is more widely available than is electrocardiographic gating. It can be used with selected thin-section CT images containing cardiac-motion artifacts without necessitating additional cost, patient dose, or scan time.
| FOOTNOTES |
|---|
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
F. Liu, C. Cuevas, A. A. Moss, O. Kolokythas, T. J. Dubinsky, and P. E. Kinahan Gas Bubble Motion Artifact in MDCT Am. J. Roentgenol., February 1, 2008; 190(2): 294 - 299. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. I. Ha, H. W. Goo, J. B. Seo, J.-W. Song, and J. S. Lee Effects of high-resolution CT of the lung using partial versus full reconstruction on motion artifacts and image noise. Am. J. Roentgenol., September 1, 2006; 187(3): 618 - 622. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| RADIOLOGY | RADIOGRAPHICS | RSNA JOURNALS ONLINE |